Data Loggers - Your Source for Temperature Data Loggers, Humidity Data Loggers, Pressure Data Loggers,Relative Humidity Data Loggers,CO data loggers,CO2 data loggers,Rainfall Gauges,Soil Moisture Content Measurement and many other data logging,measurement or recording devices

SmartReader 5 Data Logger

ACR Three Channel Data Logger: One Ambient Temperature and 2 Thermocouple Channels

ACR SmartReader 5 Data Logger: 3 Channels - Temperature and 2 Thermocouple ChannelsThe SmartReader 5 data logger is ideal for those applications that require monitoring extremely high temperatures or monitoring temperature over a wide range. The logger has one internal channel for measuring ambient temperature and two external channels for use with ACR thermocouple wire or any J, K, S or T type thermocouple wire.

Required with Smart Reader 5 Data Logger
>> TrendReader 2 Software

Required for Ethernet Connectivity
>> TrendReader 2 Software with USB Cable Kit (#TR2-USB)

NOTE: NIST Calibration is available.
Large Quantity Discounts Are Available!

Features   Specifications

Description Part Number Qty 1-9 Qty 10+  
SmartReader 5 Three Channel Temperature (thermocouple) logger SR-005
$359.00 call
Software Kit - 1 Required
TrendReader 2 Software with RS-232 Cable (IC-101) and Manual for Smart Reader series TR2-RS232
$230.00 call
TrendReader 2 Software with USB Interface Cable (IC-102) and Manual for Smart Reader series TR2-USB
$240.00 call
TrendReader 2 Software CD Only (interface cable and manual not included), for SmartReader Plus, Smart Reader, IAQ and OWL Loggers TR2-Standard
$130.00 call
Standard NIST Calibration Service
Standard NIST Temperature Calibration (1 Point) Note: 3 to 5 Day Lead Time for All Calibration Orders CAL-NIST-TEMP $49.00Temp:
Optional Accessories
USB Serial Adapter CABLE-USB232
$39.00 $35.00
ACR RS-232 Serial Interface Cable (Limited Quantity - 2 Available) LIC-101--U--
$125.00
$100.00
call
ACR RS-232 Serial Interface Cable LIC-101
$125.00 call

SmartReader 5 Logger Features

  • Monitoring performance of gas fired heating equipment
  • Verifying concrete curing temperatures
  • Process verification

Other Thermocouple Data Loggers from ACR

SmartReader 5 Logger Specifications

General Specifications
Channels Three (one internal thermistor for cold-junction compensation and two external channels for narrow range and wide range thermocouples)
Memory 32 KB (32,767 readings)
Size 107 x 74 x 22 mm. (4.2” x 2.9” x 0.9”)
Weight 110 grams (3.75 ounces)
Case Material Noryl Plastic
External Connector Removable screw-type terminal strip with Common (-) connection
Mounting Magnetic backing or locking eyelet
Operating Limits -40°C to 70°C (-40°F to 158°F) and 0 to 95%
Relative Humidity (non-condensing)
Clock Accuracy ±2 seconds per day
Battery 3.6 volt Lithium, 1 Amp-Hour
Battery Life 10 year warranty (under normal use), factory replaceable
Power Consumption 5 to 10 microamps (continuous)
Sampling Methods 1. Continuous (First-in, First-out)
2. Stop when full (Fill-then-stop)
Sampling Rates User selectable rates from 8 seconds to once every 5 days
Resolution 8 bit (1 part in 256)
Software Compatiblity Windows 3.1, 95, 98, 2000, ME, NT, XP, Vista (32-bit), Windows 7 (32-bit and 64-bit)
Software Requirements TrendReader 2 Software and Interface Cable
Certifications and Standards

- Certified to CE standard EN50082-1: 1992 (European Generic Immunity) covering ESD, RFI and EFT/B
-Meets CE standards EN55011 Class B: 1991 (European Radiated Emissions) and CISPR Class B: 1991(International Radiated Emissions)

Resistance to X-Rays - Tested for protection against a 160kV dose @ 5mA for 30
seconds (150mA-sec) @ 38 inches F.F.D. (about 100 times that of an airport x-ray machine)
- Tested for protection against Gamma Ray (equivalent to 0.137-1.38 mega volts) IR 192 - 28 curies @ 30 second exposure source to object distance 5”

Internal Temperature Sensor
Type Negative Temperature Coefficient Thermistor.
10,000 ohms @ 25°C (77°F)
Accuracy ± 0.2°C, over the range of 0 to 70°C (±0.3°F over the range of 32 to 158°F)
Range -40 to 70°C (-40 to 158°F)
Resolution 0.4°C (0.7°F) at 25°C (77°F); better than 1 °C (1.8°F) between -25°C and 70°C (-13°F and 158°F); better than 2.0°C (3.6°F) between -40°C and -25°C (-40°F and -13°F)

Thermocouples Accepted
Type Wide/Narrow Range Resolution
J Channel #2 (Wide) -50 to 600°C (-55 to 1100°F) 5°C (9°F)
Channel #1 (Narrow) -20 to 190°C (0 to 370°F) 1.4°C (2.5°F)
K Channel #2 (Wide) -100 to 900°C (-145 to 1650°F) 6.7°C (12°F)
Channel #1 (Narrow) -25 to 230°C (-10 to 440°F) 1.7°C (3°F)
S Channel #2 (Wide) 0 to 1760°C (32 to 3200°F) 23°C (41.5°F)
Channel #1 (Narrow) 0 to 960°C (32 to 1750°F) 7°C (12.6°F)
T Channel #2 (Wide) -200 to 400°C (-325 to 750°F) 6.7°C (12°F)
Channel #1 (Narrow) -35 to 200°C (-30 to 390°F) 1.7°C (3°F)
Accuracy: ± 1% of range + resolution

 

MicroDAQ.com accepts
Visa, MasterCard and American Express
Buy Data Loggers from Microdaq.com with Visa, MasterCard and American Express

All prices are in U.S. Dollars.
No U.S. sales tax