SRP-006 SmartReader Plus 6 Thermocouple Data Logger

ACR 01-0013 8 Channel Logger: One Internal Temperature, Seven Thermocouple Inputs
SKU
01-0013
$966.00
  • (7) External Thermocouple Inputs and (1) Internal Temperature Sensor
  • Accepts Type J, K, S, T or E Thermocouple Sensor Probes
  • Internal Temperature Sensor Features A Measurement Range of -40°C to 70°C (-40°F to 158°F) and ±0.2°C Accuracy
  • Powered-By a Factory Replaceable 10-Year 3.6 Volt Lithium Battery
  • 25/Second to 8 Hour User-Selectable Sampling Rate
  • Continuous (FIFO) or Stop When Full Sampling Methods
Requires ACR TrendReader 2 Software and USB Cable Kit

Due to Increased Freight Costs, All ACR Systems Orders Must Have a Minimum Order Value of $100

Special order - please allow 6-8 weeks for this item to ship.
User Manuals and Product Guides

SRP-006 ACR Systems SmartReader Plus 6 (Cat#01-0013) Thermocouple Data Logger Specifications

Channels Eight (one internal thermistor for cold-junction compensation and seven for thermocouple inputs)
Memory Available in 32KB (21,500 readings), 128KB (87,000 readings), 1.5MB (1,048,000 readings)
Size 107 x 74 x 22 mm. (4.2” x 2.9” x 0.9”)
Weight 110 grams (3.75 ounces)
Case Material Noryl Plastic
External Connector Removable screw-type terminal strip with Common (-) connection
Operating Limits -40°C to 70°C (-40°F to 158°F) and 0 to 95% Relative Humidity (non-condensing)
Clock Accuracy ±2 seconds per day
Battery 3.6 volt Lithium, 1 Amp-Hour
Battery Life 10 year warranty (under normal use), factory replaceable
Power Consumption 5 to 10 microamps (continuous)
Sampling Methods 1. Continuous (First-in, First-out – not available with sample rates faster than 8 seconds)
2. Stop when full (Fill-then-stop)
Sampling Rates User selectable rates from 25 per second to 1 reading every 8 hrs. Battery Pack (#BP-101) or Power Supply (#PS-201) required for sample rates faster than 8 seconds)
Resolution 12 bit (1 part in 4,096)
Mounting Magnetic backing or locking eyelet
OS Compatibility Windows 11, 10 and 7 (32 and 64 Bit)
Software Requirements TrendReader 2 Software and Interface Kit
Product Approvals Meets Part 15 for Digital Devices of the Code of Federal Regulations of the Federal Communications Commission (FCC). Meets radiated and conducted emission require ments for the Interference Causing Equipment Standard for Digital Apparatus of Industry Canada (ICES-003). IEC801-2, 3, 4 and EN55011 covering ESD, RFI Immunity, EFT/Burst and Radiated Emissions respectively for the Commission of the European Communities (CE)
Resistance to X-Rays Tested for protection against a 160kV dose @ 5mA for 30 seconds (150mA-sec) @ 38 inches F.F.D. (about 100 times that of an airport x-ray machine). Tested for protection against Gamma Ray (equivalent to 0.137-1.38 mega volts) IR 192 - 28 curies @ 30 second exposure source to object distance 5”

 

Thermocouples Accepted
Type Range Resolution
J -50 to 600°C (-55 to 1100°F) 0.30°C (0.55°F)
K -100 to 1150°C (-145 to 2100°F) 0.50°C (0.90°F)
S 0 to 1450°C (32 to 2642°F) 1.0°C (1.8°F)
T -200 to 400°C (-325 to 750°F) 0.40°C (0.75°F)
E -50 to 680°C (-58 to 1250°F) 0.24°C (0.44°F)
Accuracy: ± 0.5% of range + resolution

 

Internal Temperature Sensor
Type Negative Temperature Coefficient Thermistor.
10,000 ohms @ 25°C (77°F)
Accuracy ± 0.2°C, over the range of 0 to 70°C(±0.3°F over the range of 32 to 158°F)
Range -40 to 70°C (-40 to 158°F)
Resolution 0.03°C (0.05°F) at 25°C (77°F). Better than 0.07°C(0.12°F) between -25°C (-13°F) and 70°C (158°F). Better than 0.13°C (0.23°F) between -40°C (-40°F) and -25°C (-13°F)
Customer Questions
Is the software compatible with Windows 10?
Yes, it will work with Windows10
Please, mind that only logged in users can submit questions
Copyright © 2024 MicroDAQ, LLC. All Rights Reserved.