SmartReader Plus 10 Data Logger
ACR Four Channel Data Logger: One Internal Thermistor and Three
External Channels for RTD Probes
The SmartReader Plus 10 is a powerful and accurate RTD temperature logger
designed for easy monitoring and recording of temperature. Any
RTD probe or sensor with a TCR of 0.00385 can be used with this
logger.
Required with SmartReader Plus 10 Data Logger
>> TrendReader 2 Software
Required for Ethernet Connectivity
>> TrendReader 2 Software with USB Cable Kit (#TR2-USB)
>> Keyspan USB Server
>> Read the Networking and Remote Communications Guide
NOTE: Large Quantity Discounts Are Available!
Features Specifications Fast Sample Rates
| Description |
Part Number |
Qty 1-9 |
Qty 10+ |
|
| SmartReader Plus 10 128KB, 4-channels: one temp and 3 RTD's (catalog # 01-0132) |
SRP-010-128K
|
$1,050.00 |
call |
 |
| Software Kit - 1 Required |
| TrendReader 2 Software with RS-232 Cable (IC-101) and Manual for Smart Reader series (catalog # 01-0225) |
TR2-RS232
|
$230.00 |
call |
 |
| TrendReader 2 Software with USB Interface Cable (IC-102) and Manual for Smart Reader series (catalog # 01-0226) |
TR2-USB
|
$240.00 |
call |
 |
| TrendReader 2 Software CD Only (interface cable and manual not included), for SmartReader Plus, Smart Reader, IAQ and OWL Loggers (catalog # 01-0228) |
TR2-Standard
|
$130.00 |
call |
 |
| Optional Accessories |
| Fast Sampling Battery Pack (catalog # 01-0078) |
BP-101
|
$54.00 |
call |
 |
| Network/Fast Sampling Power Supply - 120V - North American Version (catalog # 01-0165) |
PS-201-NA
|
$66.00 |
$53.00 |
 |
| Network/Fast Sampling Power Supply - 220V - European Version (catalog # 01-0164) |
PS-201-EU
|
$76.00 |
$60.00 |
 |
| Extra RS-232 Interface Cable for SmartReader, SmartReader Plus and Nautilus loggers (catalog # 01-0068) |
IC-101
|
$119.00 |
call |
 |
| Keyspan USB Server for connecting USB-based Devices ( Data Logger ) over a LAN |
US-4A
|
$129.00 |
call |
 |
| USB Serial Adapter |
Cable-USB232
|
$39.00 |
$35.00 |
 |
| USB 4-Port Serial Adapter |
USA-49WG
|
$149.00 |
call |
 |
SmartReader Plus 10 Logger Features
- Data Storage Capacity of 87,000 Readings
- 4 Channels
- 10 Year Battery Life
SmartReader Plus 10 Logger Specifications
| General Specifications |
| Channels |
Four (one internal thermistor channel for
ambient temperature and three external channels for use with RTD* temperature probes)
*Any RTD with a Temperature Coefficient of Resistance (TCR) of 0.00385 can be used |
| Memory Size |
128
K (87,000 readings) |
| Size |
107 x 74 x 22 mm. (4.2” x 2.9” x
0.9”) |
| Weight |
110 grams (3.75 ounces) |
| Case Material |
Noryl Plastic |
| External Connector |
Removable screw-type terminal strip with Common (-) connection |
| Operating Limits |
-40°C to 70°C (-40°F to 158°F)
and 0 to 95%
Relative Humidity (non-condensing) |
| Clock Accuracy |
±2 seconds per day |
| Battery |
3.6 volt Lithium, 1 Amp-Hour |
| Battery Life |
10 year warranty (under normal use), factory replaceable |
| Power Consumption |
5 to 10 microamps (continuous) |
| Sampling Methods |
1. Continuous (First-in, First-out – not
available with
sample rates faster than 8 seconds)
2. Stop when full (Fill-then-stop) |
| Sampling Rates |
User selectable rates from 25 per second to
1 reading every 8 hrs. (BP-101
Battery Pack or PS-201 Power Supply required for sample
rates faster than 8 seconds) |
| Resolution |
12 bit (1 part in 4,096) |
| Mounting |
Magnetic backing or locking eyelet |
| PC Requirements |
IBM PC or 100% compatible running MS Windows
3.1,
’95, ’98, 2000, Me, NT, XP or Vista 32-bit with at least 2
MB RAM,
2 MB of hard drive disk space and one free serial port |
| Product Approvals |
Meets Part 15 for Digital Devices of the Code
of Federal
Regulations of the Federal Communications Commission
(FCC). Meets radiated and conducted emission require
ments for the Interference Causing Equipment Standard f
or Digital Apparatus of Industry Canada (ICES-003).
IEC801-2, 3, 4 and EN55011 covering ESD, RFI Immunity,
EFT/Burst and Radiated Emissions respectively for the
Commission of the European Communities (CE) |
| Resistance to X-Rays |
Tested for protection against a 160kV dose
@ 5mA for 30
seconds (150mA-sec) @ 38 inches F.F.D. (about 100 times
that of an airport x-ray machine). Tested for protection
against Gamma Ray (equivalent to 0.137-1.38 mega volts)
IR 192 - 28 curies @ 30 second exposure source to object
distance 5” |
| RTD Channels |
The following specifications are for:
4-Wire Platinum 100 Ohm @ 0° C & TCR = 0.00385
4-Wire Platinum 1000 Ohm @ 0° C & TCR = 0.00385 |
Temperature |
| Range* |
-230 to 300°C (-382 to 572°F) |
| Accuracy |
± 0.3°C (± 0.54°F) + resolution or
± 0.06% full scale (of logger) |
| Resolution |
± 0.14°C (± 0.25°F) |
Resistance (100 Ohm) |
| Range |
5 to 210 Ohm |
| Accuracy |
± 0.13 Ohm, ± 0.06 % full scale |
| Resolution |
± 0.05 Ohm |
Resistance (1000 Ohm) |
| Range |
50 to 2100 Ohm |
| Accuracy |
± 1.3 Ohm, ± 0.06% full scale |
| Resolution |
± 0.5 Ohm |
| *Temperature compensated by internal temperature sensor
for the range of -20 to 80°C (-4 to 176°F) |
Fast Sampling with Battery Pack or External 120 Volt Power Supply
For sampling rates faster than once every 8 seconds,
the Smart Reader Plus data loggers require the use of an external
120 volt power supply (#PS-201) or fast sampling battery pack (#BP-101).
The fast sampling battery pack has replaceable "AAA" batteries and 8" (20cm) cable attached, with a male stereo jack on the end. The stereo jack is plugged into the communication port on the Smart Reader Plus data logger.
The 120 Volt Power Supply is required when network length is greater than 50' , more than 3 loggers are on the network, or for sampling rates greater than 8 seconds on the SmartReader Plus Loggers.
MicroDAQ.com accepts Visa, MasterCard and American Express

All prices are in U.S. Dollars.
No U.S. sales tax
|