SmartReader Plus 6 Data Logger
ACR 8 Channel Logger: One Internal Temperature, Seven Thermocouple
Inputs
The SmartReader Plus 6 data logger is ideal for those applications that
requires
monitoring of extremely high temperatures or monitoring temperature
over a wide range. The datalogger has one internal channel for measuring
ambient temperature and seven external channels for use with thermocouple types
J, K, S, T or E.
Required with SmartReader Plus 6 Data Logger
>> TrendReader 2 Software
Required for Ethernet Connectivity
>> TrendReader 2 Software with USB Cable Kit (#TR2-USB)
>> Keyspan USB Server
>> Read the Networking and Remote Communications Guide
NOTE: NIST Calibration is available. Large Quantity Discounts Are Available!
Features Specifications Fast Sample Rates
| Description |
Part Number |
Qty 1-9 |
Qty 10+ |
|
| SmartReader Plus 6 128KB, 8-channels: one temp, seven thermocouple inputs (catalog # 01-0122) |
SRP-006-128K
|
$1,275.00 |
call |
 |
| SmartReader Plus 6 32KB, 8-channels: one temp, seven thermocouple inputs (catalog # 01-0013) |
SRP-006
|
$1,175.00 |
call |
 |
| SmartReader Plus 6 1.5MB, 8-channels: one temp, seven thermocouple inputs (catalog # 01-0151) |
SRP-006-1-5M
|
$1,625.00 |
call |
 |
| Software Kit - 1 Required |
| TrendReader 2 Software with RS-232 Cable (IC-101) and Manual for Smart Reader series (catalog # 01-0225) |
TR2-RS232
|
$230.00 |
call |
 |
| TrendReader 2 Software with USB Interface Cable (IC-102) and Manual for Smart Reader series (catalog # 01-0226) |
TR2-USB
|
$240.00 |
call |
 |
| TrendReader 2 Software CD Only (interface cable and manual not included), for SmartReader Plus, Smart Reader, IAQ and OWL Loggers (catalog # 01-0228) |
TR2-Standard
|
$130.00 |
call |
 |
| Standard NIST Calibration Service |
| Optional Accessories |
| Fast Sampling Battery Pack (catalog # 01-0078) |
BP-101
|
$54.00 |
call |
 |
| Network/Fast Sampling Power Supply - 120V - North American Version (catalog # 01-0165) |
PS-201-NA
|
$66.00 |
$53.00 |
 |
| Network/Fast Sampling Power Supply - 220V - European Version (catalog # 01-0164) |
PS-201-EU
|
$76.00 |
$60.00 |
 |
| Extra RS-232 Interface Cable for SmartReader, SmartReader Plus and Nautilus loggers (catalog # 01-0068) |
IC-101
|
$119.00 |
call |
 |
| Keyspan USB Server for connecting USB-based Devices ( Data Logger ) over a LAN |
US-4A
|
$129.00 |
call |
 |
| USB Serial Adapter |
Cable-USB232
|
$39.00 |
$35.00 |
 |
| USB 4-Port Serial Adapter |
USA-49WG
|
$149.00 |
call |
 |
SmartReader Plus 6 Logger Features
- Monitoring Performance of Gas Fired Heating Equipment
- Verifying Concrete Curing Temperatures
- Process Verification
SmartReader Plus 6 Logger Specifications
| General Specifications |
| Channels |
Eight (one internal thermistor for cold-junction compensation
and seven for thermocouple inputs) |
| Memory |
Available in 32KB (21,500 readings), 128KB (87,000 readings),
1.5MB (1,048,000 readings) |
| Size |
107 x 74 x 22 mm. (4.2” x 2.9” x 0.9”) |
| Weight |
110 grams (3.75 ounces) |
| Case Material |
Noryl Plastic |
| External Connector |
Removable screw-type terminal strip with Common (-) connection |
| Operating Limits |
-40°C to 70°C (-40°F to 158°F) and 0 to 95% Relative Humidity (non-condensing) |
| Clock Accuracy |
±2 seconds per day |
| Battery |
3.6 volt Lithium, 1 Amp-Hour |
| Battery Life |
10 year warranty (under normal use), factory
replaceable |
| Power Consumption |
5 to 10 microamps (continuous) |
| Sampling Methods |
1. Continuous (First-in, First-out – not
available with sample rates faster than 8 seconds)
2. Stop when full (Fill-then-stop) |
| Sampling Rates |
User selectable rates from 25 per second to
1 reading every 8 hrs. (BP-101
Battery Pack or PS-201 Power Supply required for sample
rates faster than 8 seconds) |
| Resolution |
12 bit (1 part in 4,096) |
| Mounting |
Magnetic backing or locking eyelet |
| PC Requirements |
IBM PC or 100% compatible running MS Windows
3.1, ’95, ’98, 2000, Me, NT, XP ot Vista 32-bit with at least 2
MB RAM, 2 MB of hard drive disk space and one free serial port |
| Software Requirements |
TrendReader 2 Software and Interface
Kit |
| Product Approvals |
Meets Part 15 for Digital Devices of the Code
of Federal Regulations of the Federal Communications Commission
(FCC). Meets radiated and conducted emission require
ments for the Interference Causing Equipment Standard for Digital Apparatus of Industry Canada (ICES-003).
IEC801-2, 3, 4 and EN55011 covering ESD, RFI Immunity,
EFT/Burst and Radiated Emissions respectively for the
Commission of the European Communities (CE) |
| Resistance to X-Rays |
Tested for protection against a 160kV dose
@ 5mA for 30 seconds (150mA-sec) @ 38 inches F.F.D. (about 100 times
that of an airport x-ray machine). Tested for protection
against Gamma Ray (equivalent to 0.137-1.38 mega volts)
IR 192 - 28 curies @ 30 second exposure source to object
distance 5” |
| Thermocouples Accepted |
| Type |
Range |
Resolution |
| J |
-50 to 600°C (-55 to 1100°F) |
0.30°C (0.55°F) |
| K |
-100 to 1150°C (-145 to 2100°F) |
0.50°C (0.90°F) |
| S |
0 to 1450°C (32 to 2642°F) |
1.0°C (1.8°F) |
| T |
-200 to 400°C (-325 to 750°F) |
0.40°C (0.75°F) |
| E |
-50 to 680°C (-58 to 1250°F) |
0.24°C (0.44°F) |
| Accuracy: ± 0.5% of range + resolution |
| Internal Temperature Sensor |
| Type |
Negative Temperature Coefficient Thermistor.
10,000 ohms @ 25°C (77°F) |
| Accuracy |
± 0.2°C, over the range of 0 to
70°C(±0.3°F over the range of 32 to 158°F) |
| Range |
-40 to 70°C (-40 to 158°F) |
| Resolution |
0.03°C (0.05°F) at 25°C (77°F).
Better than 0.07°C(0.12°F) between -25°C (-13°F) and 70°C
(158°F). Better than 0.13°C (0.23°F) between -40°C
(-40°F) and -25°C (-13°F) |
Fast Sampling with Battery Pack or External 120 Volt Power Supply
For sampling rates faster than once every 8 seconds,
the Smart Reader Plus data loggers require the use of an external
120 volt power supply (#PS-201) or fast sampling battery pack (#BP-101).
The fast sampling battery pack has replaceable "AAA" batteries and 8" (20cm) cable attached, with a male stereo jack on the end. The stereo jack is plugged into the communication port on the Smart Reader Plus data logger.
The 120 Volt Power Supply is required when network length is greater than 50' , more than 3 loggers are on the network, or for sampling rates greater than 8 seconds on the SmartReader Plus Loggers.
MicroDAQ.com accepts Visa, MasterCard and American Express

All prices are in U.S. Dollars.
No U.S. sales tax
|