| Data Channels |
Single-Ended: 16 Analog Channels
Differential: 8 Analog Channels (Single-Ended or Differential Available, Not Both)
Analog to Digital Converter: 2
3 or 4 Wire Pt100/1000: 4
Pulse: 4 (2 x 64kHz & 2 x 100Hz)
Event (Digital): 8 State Inputs or 1 8-Bit Binary
High Voltage: 2
Temperature: 1 (Internal) |
| Analog Inputs |
Accuracy (Voltage & Resis.): ±(0.05% + 0.025% Range)
Common Mode Rejection: 100dB
Input Impedance: > 1M Ω
Linearity: 0.015%
Series Mode Line Rejection: 50/60Hz 100dB |
| Analog to Digital Conversion |
Type: Sigma - Delta
Resolution: 24 Bit
Sampling Rate: up to 10, 20* or 100* Readings per Second
* With Mains rejection Off. |
| Alarm Outputs |
4 (Open Drain FET - 18.0 Volts - 0.1 Amps) |
| Power Output |
External Devices: Regulated 5 Volts DC at 50 mA or Logger Supply Voltage (100mA) |
| Time and Date |
Built-In Clock in 3 Formats |
| Scaling Data |
Displays Readings in Preferred Engineering Units |
| Data Storage Capacity |
Internal: 16MB Storing up to 14 Million Samples/Readings
External: 1 GB Removable SD Card |
| DC Voltage Ranges |
-0.075 to 0.075 Volts -0.15 to 0.15 Volts -0.3 to 0.3 Volts -0.6 to 0.6 Volts -0.6 to 1.2 Volts -0.6 to 2.4 Volts
-3.0 to 3.0 Volts -6.0 to 6.0 Volts
-6.0 to 12.0 Volts
-6.0 to 25.0 Volts
4.0 to 20.0 Volts*
4.0 to 40.0 Volts*
4.0 to 60.0 Volts*
* High Voltage (Max of 2 Can be Selected) |
| Current Range |
Differential: 4 to 20mA, -30mA to 30.0mA
(Requires External 10Ω Shunt) |
| Thermocouple Range |
J-Type: -200°C to 1200°C (-328°F to 2192°F)
K-Type: -200°C to 1372°C (-328°F to 2501°F)
N-Type: -200°C to 1300°C (-328°F to 2372°F)
R-Type: -50°C to 1768°C (-58°F to 3214°F)
S-Type: -50°C to 1768°C (-58°F to 3214°F)
T-Type: -200°C to 400°C (-328°F to 752°F)
(Differential and Single Ended) |
| Resistance Range |
3-Wire and 4-Wire: 0 to 500Ω, 0 to 4,000Ω |
| Thermistor Range |
U & Y-Type: -50°C to 150°C (-58°F to 302°F) |
Pt100/1000 Range
(3 or 4-Wire) |
-200°C to 850°C (-328°F to 1562°F) |
| Internal Reference Temperature |
-50°C to 150°C |
| Pulse Count Range |
0 to 100Hz (2 Input)
0 to 64kHz (2 Input)
0 to 16,000,000 Count |
| Digital State/Event Ranges |
8 State Inputs or 1 x 8 Bit Binary |
| Digital/Alarm Outputs |
2 Open Drain FETs, 18 Volts, 0.1 Amp |
| Calculated Channels |
Up to 16 Virtual Channels Derived from Physical Input Channels |
| Resolution |
Up to 6 Significant Digits |
| Communication |
Standard: RS-232 (Auto Bauding to 115,200 Baud)
USB: 1.1 and 2.0 Compatible
Other: GSM, Wi-Fi and PSTN Modems (External Attachments) |
| Wireless Communication |
Wireless Ethernet (Wi-Fi): 802.11b, 2.4GHz, 1 to 14 Channels
Security: Open, WEP (64 or 128bit Encryption), WPA or WPA2/802.11i
Network: Infrastructure Only with Specified SSID, or any Network with no SSID
Note: External Power Supply Required for Wi-Fi |
| Power Supply |
Internal: 6 AA Alkaline Batteries
External: 10 to 18 Volts DC Power Supply
NOTE: Reverse Polarity and Over-Voltage Protected |
| Power Consumption |
Sleep Mode: 600 µA
Logging: 40 to 80 mA |
| Display/Keypad |
2 Line x 20 Characters LCD Display |
| Display Functions |
Battery State and External Power Indicator Keypad Lock Arm/Disarm/Pause/Continue Meter Any Channel or Alarm Select up to 6 Pre-Stored Setups Status/Diagnostics/Memory/Time and Date Download to MicroSD Card |
| Enclosure |
ABS Material |
| Operating Modes |
>> Stop When Full >> Overwrite |
| Operating Range |
Temperature: -30°C to 65°C
Humidity: 0 to 90% at 40°C non-condensing |
| Standards Compliance |
CE |
| Weight |
1.2 kilograms |
| Dimensions |
23.5 cm x 17.5 cm x 0.55 cm (175mm x 135mm x 55mm) |