Data Loggers - Your Source for Temperature Data Loggers, Humidity Data Loggers, Pressure Data Loggers,Relative Humidity Data Loggers,CO data loggers,CO2 data loggers,Rainfall Gauges,Soil Moisture Content Measurement and many other data logging,measurement or recording devices

GL1100 Hard Disk Data Logger

Graphtec Logger with 40 GB Hard Drive and Sampling Rate up to 40 MS/s per Channel

The GL1100 Hard Disk Logger provides high speed and simultaneous sampling of voltage, voltage/temperature, strain and frequency amplifiers. The data logger features high speed sampling up to 40 MS/Second per Channel, Web and FTP Server, Internal 40GB hard drive, Built-in Ethernet and USB Interfaces, PCM drive slot, Internal 2 MS memory and accepts 4 or 8 GL11 Series preamps.

2 Versions of GL1100 Logger Available
>> 8-Channel (#GL1100-8)
>> 16-Channel (#GL1100-16)

PreAmps (Amplifiers) Available
>> 2-Channel Voltage Module (#GL11-HSV)
>> 2-Channel Multi-Input Module (#GL11-M)
>> 2-Channel DC Strain Module (#GL11-DCB)

Included With Each Logger
>> User Manual and Software

Features   Specifications   Amplifiers

NOTE: Large Quantity Discounts available!

The GL1100 Data Logger Has Been Discontinued by the Manufacturer.
Contact Sales @ 1-877-275-9606 for a Replacement Product.

Hard Disk Logger Features

  • Available in 8 or 16-Channel Input Versions
  • Up to 40 MS/s Sampling per Channel
  • Allows combined use of Voltage, Voltage/Temp, Strain and Frequency Amplilfiers
  • Built-in Web and FTP Server to share and transfer data
  • Optional 8.4" Color TFT LCD Display and Printer
  • Includes Windows software for setup, control, data transfer, data conversion to ASCII and Excel formats, and on-screen data review with cursors
  • Requires at least 1 PreAmplifier Plug-in to Operate

Hard Disk Logger Specifications

Analog Input GL1100-8: 8 Channels (4 Slots)
GL1100-16: 16 Channels (8 Slots)
Logic Input GL1100-8: 8 Channels
GL1100-16: 16 Channels
Internal Memory Storage GL1100-8: 2 Mword/channel
GL1100-16: 2 Mword/channel
Hard Drive Storage 40 GB
Synchronize Function N/A
PC Interface PCMCIA Slot (Type II), Ethernet LAN or USB
Operating Environment Temperature: 0 to 40C
Humidity: 30 to 80% RH
Rated Power Supply GL1100-8: AC Adapter (100V to 240V AC)
GL1100-16: N/A
Power Consumption GL1100-8: 90 VA Max
GL1100-16: 135 VA Max
External Dimensions GL1100-8: 11.81" x 8.74" x 2.24"
GL1100-16: 11.81" x 8.74" x 4.48"
Excludes Rubber Boot
Weight GL1100-8: 2.4kg (includes 4 amplifiers)
GL1100-16: 4.2kg (includes 8 amplifiers)
Optional AC Adapter Excluded

Hard Disk Logger Amplifiers

2-Channel Voltage Preamp Module

  • Floating Ground with Unbalanced Load Input (Independent Channels)
  • 200mV to 200 Volt Full-Scale
  • 25 ns Sample Interval (40MHz) @ 12-bit Resolution
  • Frequency Response: DC to 10 MHz
  • Input Filter: 5, 50, 500, 5k, 50k, 500k (Hz)

2-Channel Multi-Input Preamp Module

  • Floating Ground with Unbalanced Load Input (Independent Channels)
  • Voltage Input: 20 MV to 500V Full-Scale
  • Thermocouple Inputs: J, K, T, R, E, B
  • 10 us Sample Interval (100kHz) @ 16-Bit Resolution
  • Frequency Response: DC to 20 kHz
  • Input Filter: 1.5, 5, 10, 30, 50, 500, 5k (Hz)

2-Channel DC Strain Preamp Module

  • Floating Ground with Unbalanced Load Input (Independent Channels)
  • 1000 to 20,000 x 10-6 Strain Full-Scale
  • 10 us Sample Interval (100kHz) @ 16-Bit Resolution
  • Frequency Response: DC to 20 kHz
  • Input Filter: 1.5, 5, 30, 100, 300, 1k (Hz)

 

MicroDAQ.com accepts
Visa, MasterCard and American Express
Buy Data Loggers from Microdaq.com with Visa, MasterCard and American Express

All prices are in U.S. Dollars.
No U.S. sales tax