SRP-010 SmartReader Plus 10 RTD Temperature Data Logger
ACR Four Channel Data Logger: 1 Internal Thermistor and 3 External Channels for RTD Probes
SKU
01-0132
$1,262.00
- (3) External RTD Inputs and (1) Internal Thermistor Temperature Sensor
- Accepts Any 4-Wire Platinum 100/1000 Ohm RTD Sensor Probe with a TCR of 0.00385
- Internal Temperature Sensor Features A Measurement Range of -40°C to 70°C (-40°F to 158°F) and ±0.2°C Accuracy
- Powered-By a Factory Replaceable 10-Year 3.6 Volt Lithium Battery
- 25/Second to 8 Hour User-Selectable Sampling Rate
- Continuous (FIFO) or Stop When Full Sampling Methods
Due to Increased Freight Costs, All ACR Systems Orders Must Have a Minimum Order Value of $100
Special order - please allow 6-8 weeks for this item to ship.
SRP-010 ACR Systems SmartReader Plus 10 (Cat#01-0132) RTD Temperature Data Logger Specifications
Channels | Four (one internal thermistor channel for ambient temperature and three external channels for use with RTD* temperature probes) *Any RTD with a Temperature Coefficient of Resistance (TCR) of 0.00385 can be used |
|
Memory Size | 128 K (87,000 readings) | |
Size | 107 x 74 x 22 mm. (4.2” x 2.9” x 0.9”) | |
Weight | 110 grams (3.75 ounces) | |
Case Material | Noryl Plastic | |
External Connector | Removable screw-type terminal strip with Common (-) connection | |
Operating Limits | -40°C to 70°C (-40°F to 158°F) and 0 to 95% RH (non-condensing) | |
Clock Accuracy | ±2 seconds per day | |
Battery | 3.6 volt Lithium, 1 Amp-Hour | |
Battery Life | 10 year warranty (under normal use), factory replaceable | |
Power Consumption | 5 to 10 microamps (continuous) | |
Sampling Methods | 1. Continuous (First-in, First-out – not available with sample rates faster than 8 seconds) 2. Stop when full (Fill-then-stop) |
|
Sampling Rates | User selectable rates from 25 per second to 1 reading every 8 hrs. (BP-101 Battery Pack or PS-201 Power Supply required for sample rates faster than 8 seconds) | |
Resolution | 12 bit (1 part in 4,096) | |
Mounting | Magnetic backing or locking eyelet | |
OS Compatibility | Windows 11, 10 and 7 (32 and 64 Bit) | |
Product Approvals | Meets Part 15 for Digital Devices of the Code of Federal Regulations of the Federal Communications Commission (FCC). Meets radiated and conducted emission require ments for the Interference Causing Equipment Standard f or Digital Apparatus of Industry Canada (ICES-003). IEC801-2, 3, 4 and EN55011 covering ESD, RFI Immunity, EFT/Burst and Radiated Emissions respectively for the Commission of the European Communities (CE) |
|
Resistance to X-Rays | Tested for protection against a 160kV dose @ 5mA for 30 seconds (150mA-sec) @ 38 inches F.F.D. (about 100 times that of an airport x-ray machine). Tested for protection against Gamma Ray (equivalent to 0.137-1.38 mega volts) IR 192 - 28 curies @ 30 second exposure source to object distance 5” |
RTD Channels | |
---|---|
The following specifications are for: 4-Wire Platinum 100 Ohm @ 0° C & TCR = 0.00385 4-Wire Platinum 1000 Ohm @ 0° C & TCR = 0.00385 |
|
Temperature
|
|
Range* | -230 to 300°C (-382 to 572°F) |
Accuracy | ± 0.3°C (± 0.54°F) + resolution or ± 0.06% full scale (of logger) |
Resolution | ± 0.14°C (± 0.25°F) |
Resistance (100 Ohm)
|
|
Range | 5 to 210 Ohm |
Accuracy | ± 0.13 Ohm, ± 0.06 % full scale |
Resolution | ± 0.05 Ohm |
Resistance (1000 Ohm)
|
|
Range | 50 to 2100 Ohm |
Accuracy | ± 1.3 Ohm, ± 0.06% full scale |
Resolution | ± 0.5 Ohm |
*Temperature compensated by internal temperature sensor for the range of -20 to 80°C (-4 to 176°F) |
Customer Questions