Lascar USB Thermistor Vaccination Monitoring Kit

  • Lascar USB Thermistor Vaccination Monitoring Kit
  • Lascar USB Thermistor Vaccination Monitoring Kit


Lascar USB Thermistor Vaccination Monitoring Kit

Part Number: EL-USB-VAC

Ask a question (4)

Be the first to review this product


  • Monitor Vaccine Temperatures Using External Thermistor Probe

  • Stores 32,000 Temperature Readings onto Non-Volatile Internal Memory

  • Powered by User-Replaceable 1/2 AA Lithium Battery

  • 2 User-Programmable Alarm Thresholds with Visual LED Indicators

  • USB Interface for Set-up and Data Download

  • Includes a NIST Traceable Calibration Certificate, Compliant to ISO/IEC 17025:2005

  • Unless otherwise stated by the customer the calibration interval will default to the manufacturers recommendation of 1-Year

  • Data Loggers are Calibrated for Both Fridge (5°C) and Freezer (-15°C) Use

  • Can Be Used as Either Primary or Backup Data Logger

  • Meets 2018 CDC & VFC (Vaccines for Children) Program Recommendations for Safe Vaccine Storage & Handling


Includes NIST Traceable Calibration Certificate, 1 Meter Thermistor Sensor Probe attached to Glycol Bottle, Protective Cap and 1/2 AA 3.6 Volt Lithium Battery. PC Compatible Software Available for Download.

20% Restocking Fee if Returned for Order Error.

$115.00

Lascar USB Vaccine Monitoring Kit Specifications

Temperature Measurement Range: -40°C to 125°C (-40°F to 257°F)
Resolution: 0.5°C (1.0°F)
Accuracy: ±0.5°C (±1.0°F)
Response Time: 40 Minutes to 90%
External Sensor and Glycol Bottle Sensor Type: 75mm Stainless Steel Capped Probe
Cable Length: 1.0 Meter (3ft)
Data Storage Capacity 32,000 Samples
Sampling Rate 1 Second to 12 Hours
Start Modes Immediate, Delayed and Push-to-Start
Operating Range Temperature: -35°C to 80°C (-31°F to 176°F)
Note: Data Logger Should Not be Subjected to Temperatures Outside Operating Range
Battery User Replaceable 1/2AA 3.6V Lithium
Battery Life Typically 2 Years
OS Compatibility Windows 10, 8 and 7 (32-bit and 64-Bit Versions)
LED Green: Recording and Logging
Red: Alarm Thresholds Met
Orange: Low Battery or Logger Data Capacity full
Display Functions Recent Logged Temperature Reading
Maximum Logged Temperature
Minimum Logged Temperature
Alarm Visual - Blinking Red LED
Standards Compliance RoHS
Weight 1.52 oz
Dimensions 13.45 cm x 2.12 cm x 2.39 cm (5.29" x 0.83" x 0.94")

Customer Reviews

Write Your Own Review

You're reviewing: Lascar USB Thermistor Vaccination Monitoring Kit

How do you rate this product? *

  1 star 2 stars 3 stars 4 stars 5 stars
Price
Quality
Value

Questions

  • Answer:

    While technically you as the customer can request any calibration interval you want, the VFC typically won't accept going with any interval that is greater than 2 years or whichever the manufacturer recommends.  For this particular logger the manufacturer doesn't recommend going longer than a year.  LogTag recommends up to 2 years and Berlinger and InTemp suggest up to 3 years.  Keep in mind that when a logger is calibrated and is found to be accurate that is only at the point that the calibration was performed.  While the accuracy of most temperature loggers hold pretty accurate it's not uncommon for issues to pop up that cause it to drift out of spec.  Should this happen and you don't have it calibrated/tested regularly you might not find out about an issue until it's too late.

    All the vaccine monitoring loggers we sell with the exception of the Fridge-Tag2L series from Berlinger can be shipped back to us for recalibration.  The cost for recalibration is much less expensive than buying a new logger, typically it's under $35 per logger.

Leave a Reply

  • Reload captcha
  • Answer:

    As the customer you can request any interval you would like however the manufacturer of this product doesn't recommend anything more than a one year interval.  If you order online you can state the interval you wish us to list in the order notes during checkout.  

Leave a Reply

  • Reload captcha
  • Answer:

    Yes, this logger meets the current requirements of the VFC program.

    The logger uses a 1/2 AA battery (included) and has a typical battery life of up to 1-2 years. 

Leave a Reply

  • Reload captcha
  • Answer:

    Generally speaking we don't recommend keeping a logger in a freezer unless it's for short term testing.  This is because the cold temperatures greatly reduce the battery life and make the LCD display (if equipped) not readable.  To use a logger long term at -40°C we would recommend looking at a logger rated for dry ice applications, however even these loggers will have a relatively short battery life and most (if not all) don't have displays. 

    If you are using this to meet a requirement of the VFC program they require you to use a logger with an external probe, display, and glycol buffer for the sensor.  There is no way to meet this requirement with a logger kept inside the freezer. 

Leave a Reply