Onset Differential Air Pressure Transducer Sensor

SKU
T-VER-PX3UL
$279.00
  • Measures Air Pressure or Air Velocity
  • 0.1,0.25, 0.5,1.0" WC FS (UniDirectional) and ±0.1", ±0.25", ±0.50", and ±1.0" WC FS (BiDirectional)
  • 2-Wire, Loop-Powered 4-20mA (DC Only, Clipped and Capped) or 3-Wire 0-5V/0-10V Output Field-Selectable
  • Signed 3-1/2 digit LCD Display
  • Accuracy of ±1% F.S. of Selected Range
Special order - please be advised that due to supply shortages, some items may require 6 weeks for shipment
User Manuals and Product Guides
  • Media Compatibility Dry Air or Inert Gas
  • Input Power: 12-30VDC, or 24VAC nominal
  • Output Field: 2-wire, loop-powered 4-20mA (DC only, clipped and capped) or 3-wire 0-5V/0-10V
  • Unidirectional Pressure Ranges: 0.1/0.25/0.5/1.0” W.C. F.S. (25 pa/50 pa/100 pa/250 pa, F.S.), Switch Selectable
  • Bidirectional Pressure Ranges: ±0.1/±0.25/±0.5/±1.0” W.C. F.S. (±25 pa/±50 pa/±100 pa/±250 pa, F.S.), Switch Selectable
  • Response Time: T95 in 20 sec (Standard), T95 in 2 sec (Fast), Switch Selectable
  • Mode: Unidirectional or bidirectional, switch selectable
  • Display: Signed 3-1/2 digit LCD, indicates pressure, over range indicator
  • Proof Pressure: 3 psid (20.6kPa)
  • Burst Pressure: 5 psid (34.5kPa)
  • Accuracy: ±1% F.S. of selected range (combined linearity and hysteresis)
  • Temperature Effect: 1" 0.05%/°C; 10"
  • Zero: Drift (1-year) 1" 2.0% max.
  • Zero Adjust: Pushbutton auto-zero and digital input (2-pos terminal block)
  • Operating Environment: 0°- 60°C (32° to 140°F); 0 to 90% RH non-condensing
  • Fittings: Brass barb; 0.24” (6.1mm) o.d.
  • UL 94 V-O Fire Retardant ABS
Customer Questions
No Questions
Please, mind that only logged in users can submit questions