Onset HOBO® UX120-006M 4-Channel Analog Data Logger Specifications
Data Channels | (4) External Inputs for Multiple Parameters |
Analog Inputs | Number of Inputs: 4 Measurement Range: 0 to 20.1mA (w/ CABLE-4-20MA) 0 to 2.5V (w/ CABLE-2.5-STEREO) 0 to 5V (w/ CABLE-ADAPT5) 0 to 10V (w/ CABLE-ADAPT10) 0 to 24V(w/ CABLE-ADAPT24) Accuracy: 4-20mA ±0.001mA ±0.2% of Reading (w/ CABLE-4-20MA) ±0.1 mV ±0.01% of Reading (w/ CABLE-2.5-STEREO) ±0.2 mV ±0.03% of Reading (w/ CABLE-ADAPT5) ±0.4 mV ±0.03% of Reading (w/ CABLE-ADAPT10) ±0.1 mV ±0.03% of Reading (w/ CABLE-ADAPT24) Resolution: 0.3µA 4-20mA (w/CABLE-4-20MA) 40µV (w/CABLE-2.5-STEREO) 80µV (w/CABLE-ADAPT5) 160µV (w/CABLE-ADAPT10) 384µV (w/CABLE-ADAPT24) |
Data Storage Capacity | 1.9 Million Temperature Readings |
Sampling Rate | 1 Second to 18 Hours (User-Specified) |
Display | Type: LCD Refresh Rate: 15 Seconds Note: Display is Visible from 0°C to 50°C (32°F to 158°F) Display May React Slowly or Go Blank in Temperatures Outside this Range. |
Power Source | (2) AAA 1.5 Volt Alkaline or Lithium Batteries |
Battery Life | 1 Year @ 1 Minute Sampling Rate (Dependant Upon Environment and Operating Range) |
Alarms | Visual High and Low Alarm Thresholds Displayed on LCD and Set via Software |
Time Accuracy | ±1 Minute per Month @ 25°C (32°F to 122°F) |
Resolution | 16 Bit |
Communications | USB Connection to Setup Data Logger and Download Data |
Logging Modes | Normal, Burst or Statistics |
Data Storage Modes | Stop When Full or Wrap/Overwrite When Full (FIFO) |
Start Modes | Immediately, Push Button, Date & Time or Next Interval |
Restart Mode | Push Button |
Stop Modes | When Data Storage Full, Push Button or Date & Time |
Software Compatibility | HOBOware and HOBOware PRO Version 3.5+ |
OS Compatibility | Windows: Windows 10, 8.1, 8 and 7 Mac: OS X 10.9.x, 10.10.x and 10.11.x |
Operating Range | Temperature: -20°C to 70°C (-4°F to 158°F) Humidity: 0 to 95% RH (non-condensing) Launch/Readout: 0°C to 50°C (32°F to 122°F) per USB Specification |
Standards Compliance | CE, IP50 |
Weight | 107.5 grams (3.79 oz.) |
Dimensions | 10.8 cm x 5.41 cm x 2.54 cm (4.25" x 2.13" x 1.0") |
You may also be interested in the following product(s)
Onset HOBO 4-20 mA Input Cable - CABLE-4-20MA
|
$25.00
|
|
CTV-A 0-20 Amp split-core AC Current Sensor
|
$95.00
|
|
CTV-B 0-50 Amp split-core AC Current Sensor
|
$95.00
|
|
CTV-C 0-100 Amp Split-Core AC Current Sensor
|
$95.00
|
|
CTV-D 0-200 Amp Split-Core AC Current Sensor
|
$120.00
|
|
ConLab AC Voltage Transmitter Sensors
Requires 4-20 mA Cable (#CABLE-4-20MA) NOTE: These Sensors Tie Directly into the Line And Should be Used Only By Qualified Personnel! Learn More |
$139.00
Voltage Range |
|
CTV-E 0-600 Amp Split-Core AC Current Sensor
|
$120.00
|
|
Air Velocity and Temperature Remote Head Sensor
|
$400.00
Measurement Range |
|
Onset HOBO AC-SENS-1 AC Power Adapter
Compatible with Onset HOBO U12, U30 and ZW Series Data Loggers Learn More |
$29.00
|
|
Onset 0 to 5 Volts DC External Input Cable
|
$12.00
|
|
Onset 0 to 10 Volts DC External Input Cable
|
$12.00
|
|
Onset 0 to 24 Volts DC External Input Cable
|
$12.00
|
|
Telaire 7001 Carbon Dioxide (CO2) Monitor
|
$529.00
Purchase Options |
|
Telaire CO2 Monitor to Data Logger Cable
|
$29.00
|
|
Compressed Air FlowMeter
Includes 110 Volt DC Power Adapter w/ 20 ft. Cable Learn More |
$629.00
|
|
Compressed Air Flow Meter
Includes 110 Volt DC Power Adapter w/ 20 ft. Cable Learn More |
$495.00
|
|
Onset Split Core Bi-Directional Current Transducer
|
$245.00
|
|
Onset 0 to 2.5 Volts DC Voltage Input Cable - CABLE-2.5-STEREO
|
$10.00
|
|
Ashcroft Gauge Pressure Sensor
Requires Analog Port Selection During U30 or RX3000 Series System Configuration and Use of a S-FS-CVIA when using the H22-001 Data Logger. When using a U12 or UX120 Series Data Loggers, this Sensor Requires a 0-5 VDC Analog Input Cable (CABLE-ADAP5) and External Power Provided by an AC Adapter (AC-SENS-1). Learn More |
$199.00
Gauge Pressure Range |
|
Veris 480 V, 100 AMP Kilowatt Transducer
Includes 3 100 AMP Split-Core CTs Learn More |
$845.00
|
|
Onset TMCx-HD Series Temperature Sensor
Cable Available in 1 ft, 6 ft, 20 ft and 50 ft Lengths Learn More |
$35.00
Cable Length |
|
Onset Stainless Steel Temperature Probe
Includes 1.8 Meter (6ft) Cable Learn More |
$87.00
|
|
Onset Air Temperature Sensor
Includes Tie Wraps, Screw and Thermal Grease Learn More |
$39.00
|
|
Volatile Organic Compound Sensor
Includes User Manual Learn More |
$3,999.00
|
Customer Reviews
- An Indispensible Tool For Troubleshooting Review by Arfur Smithers
-
Price Quality Value
The only caveat I would add is to make sure you have a MINI USB cable. It does not come with one. Nowadays it is very easy to find MICRO USB cables -- but MINI cables are becoming harder to find. Without that, you can't configure or use this device. Order one when you order this data logger. The price is negligible, compared to the time you will waste trying to track one down. Ideally, they would include one, but they don't. (Posted on 4/24/2016)
Write Your Own Review
Manuals and Resources
Questions
-
Answer:
No, this logger doesn't come with a calibration certificate. If you require a calibration certificate you should add it as an option. This can be selected right on the product page. Once selected you will be offered a drop down menu to choose the calibration points.
-
Answer:
Very good question. Most customers get along fine with the free version of HOBOware. The pro version has some addition capabilities and is required on certain Onset products (not this one).
Here is a comparison between the two so you can review the differences - https://www.microdaq.com/hoboware-software-comparison
-
Answer:
The total number of readings this logger can record on its internal memory is 1.9 million readings.
This is the total number or data values for all four channels so if you were recording on all four channels you could record 475k readings per channel.
The data can be transferred to a computer after the recording is completed but you can't stream the data to have it record on the computer for infinite recording.
-
Answer:
While the combination of this logger and the 0-2.5v input cable will allow you to record the presence of voltage this logger doesn't have a fast enough sample rate to capture events.
For a 4 channel event logger we would recommend the HOBO UX120-017M found here - https://www.microdaq.com/onset-hobo-ux120-pulse-event-runtime-data-logger.php