MicroDAQ.com
Toll Free: 1-877-275-9606
(603) 746-5524
$5.99 Flat Fee Economy Shipping on Most Items Shipped in the Continental USA

You have no items in your shopping cart.

Subtotal: $0.00

Lascar EL-USB-3 USB Voltage Data Logger



USB Voltage Data Logger

Lascar Direct Connect Data Recorder Measuring and Recording 0 to 30 Volts DC

Part Number: EL-USB-3

Ask a question (3)

Be the first to review this product


  • Measure Voltage Readings Over a 0 to 30V DC Range

  • USB Interface for Set-up and Data Download

  • User-Programmable Alarm Thresholds

  • Status Indication via Red and Green LEDs

  • User-Replaceable Internal Lithium Battery with 3-Year Typical Battery Life


Includes Measurement Leads with Terminated Crocodile Clips and 1/2 AA 3.6V Lithium Battery. PC Compatible Software Available As A Download.

$75.00
  • Buy 10 for $71.25 each and save 5%
  • Buy 20 for $67.50 each and save 10%

All Calibration Orders Have a 5 to 7 Business Day Lead Time
1 Year Calibration Due Date Unless Otherwise Specified

$75.00

EasyLog EL-USB-3 USB Voltage Logger Specifications

Data Storage Capacity 32,000 Samples/Readings
Measurement Range 0 to 30 VDC
Internal Resolution 50 mV DC
Accuracy (Overall Error) ±1%
Logging Rate 1s, 10s, 1m, 5m, 30m, 1hr, 6hr, 12hr
Logger Operating Range -25 to 80°C(-13 to 176°F)
Power Source 1/2AA 3.6V Lithium Battery
Battery Life 3 Year Typical Life
OS Compatibility Windows 10, 8 and 7 (32-bit and 64-Bit Versions)
Standards Compliance RoHS
*Depending on sample rate, ambient temperature and use of alarm LEDs

You may also be interested in the following product(s)

BTG-RK Locking Protective Plastic Wall Cover

BTG-RK Locking Protective Plastic Wall Cover

  • Clear Plastic
  • Vertical or Horizontal Mounting
  • Dimensions: 6 3/8" x 3 1/2" x 3"

Includes Locking Key and Mounting Hardware

Learn More
$11.00

Customer Reviews

Write Your Own Review

You're reviewing: USB Voltage Data Logger

How do you rate this product? *

  1 star 2 stars 3 stars 4 stars 5 stars
Price
Quality
Value

Questions

  • Answer:

    This type of application isn't possible with this logger due to a few reasons.  First, the logger doesn't have a real time output so it won't log while connected to a PC.  The USB jack is only used to plug the logger in for configuration and to download data after its been recorded.  The other reason is that the software can only connect to one logger at a time.  You can use two loggers if you wanted and then open both data files within the software to compare the two. 

Leave a Reply

  • Reload captcha
  • Answer:

    Yes, the software will work on both Windows 7 and 10.  

Leave a Reply

  • Reload captcha
  • Answer:

    The input impedance for the EL-USB-3 is 60.7 kilohm and 5pF in parallel.

Leave a Reply

  • Reload captcha

Ask Your Question

  • Reload captcha

* Required Fields