5396-0511 Track-It USB Voltage Data Logger

Monarch Instrument Voltage and Temperature Data Logger with Direct USB Connection
SKU
5396-0511
In Stock
$144.00
  • Measures and Records Temperature (-20°C to 70°C) and DC Voltage (0 to 500 milliVolts)
  • Stores a Total of 64,000 Samples/Readings (32,000 Temperature and 32,000 Voltage)
  • User-Configurable Alarms and LED Indicators
  • Direct USB Interface for Set-up and Data Download

Includes Lithium Coin Cell Battery and Software is Available as a Free Download

Overview of the Monarch Track-It 5396-0511 Voltage Data Logger

The Monarch Instruments 5396-0511 Track-It voltage data logger measures 0 to 500 mV input signals and temperature. The removable, swappable input modules make this an extremely versatile series of data loggers. Different input modules can be purchased and added to the standard Track-It temperature and humidity loggers allowing you to measure and record temperature, humidity, as well as 500 mV (this model), 1v, 5v, 10v or 0-20 mA inputs. The 5396-0511 featured here includes a temperature logger and the 0 to 500 mV input module. The logger can measure a temperature range of -20°C to +70°C (-4°F to +158°F) with an accuracy of +/-0.5°C. When used with the voltage input module a total of 32k readings can be recorded before downloading the logger when recording two channels or 64k readings when recording one channel. Once connected the logger is configured and data is downloaded with the easy to use but powerful Track-It Software which is available as a free download. The logger can be configured to start and stop at a predetermined time and date or only when an alarm condition is sensed. The sample rate is programable anywhere from once every 2 seconds to one sample every 24 hours.

All Track-It series USB data loggers are compatible with the On-The-Go (OTG) cable (sold separately) which allows you to download data in the field using an Android device running the free Track-It Transport app.

Software Downloads
zip
(Size: 111.45 MB)
Version 1.0.24 - Compatible with All Monarch Temperature, Humidity and Pressure Data Loggers
User Manuals and Product Guides
pdf
(Size: 0.26 MB)
Details on the Connection of the Track-It External Input Modules

Monarch Instrument 5396-0511 USB Voltage Track-It Data Logger Specifications

Data Storage Capacity 64,000 Total Samples/Readings
(NOTE: If Attaching an Analog Input Module, Only 32,000 Temperature Readings Available)
Sampling Rate 2 Seconds to 24 Hours (User-Selectable via Software)
Voltage Range: 0 to 500 mV DC
Accuracy: ±0.5% of Full Scale
Resolution: ±0.1 mV
Temperature Range: -20°C to 70°C (-4°F to 158°F)
Accuracy (0°C to 50°C): ±0.5°C
Accuracy (20°C to 70°C): ±2.0°C
Resolution: 0.6°C (0.11°F)
Repeatability: ±0.1°C (±0.2°F)
Time Accuracy ±1 Minute per Year
Triggers Number: 2 Independant Triggers
Modes: Instanteous, Button Control, On Alarm, Time and Date, Day of Week
Battery 3-Volt CR-2032 Lithium Battery (User Replaceable)
Battery Life Typically 1 Year @ 1 Minute Sampling Rate
Operating Modes >> Continuous
>> Start and Stop Using Buttons on Front of Logger
>> Start and Stop at Predetermined Date/Time
>> Start When Alarm Condition is Sensed
Display Type: 4 Digit LCD
Functions: Temperature, Engineering Units, Alarms, Record Mode, Battery Condition, Time/Data Stamp and Memory
LED User Programmable: Alarm Indicator, Recordings and Triggers
Button User Programmable: View Display, Record On/Off, Reset Alarms, Reset Timer
Alarms (2) User Programmable Visual Alarms. High or Low
Communications Interface Direct USB Connection
Software Track-It Software Required (Available as a Download)
OS Compatibility Windows 11, 10 and 7 (32 and 64 Bit)
Standards Compliance CE
Dimensions 103.2mm x 16.7mm x 25.4mm (4.06" x 0.66" x 1.00")
Customer Questions
Is the software compatible with Windows 10?
Yes, it will work with Windows10
On the Track-It USB Voltage Data Logger, one channel is an analog voltage input, what type of input is the second channel if internal temperature is not selected. is it type K thermocouple, analog voltage, RTD, thermister or not used? The Specifications or the manual do not seem to be clear on this point.
There are two inputs for varying voltage levels however they are not isolated.  Deselecting the temp sensor only gives you additional memory.  It does not allow other input types.  Both analog inputs are the same type of whichever voltage...
Is this device does the measurement of voltage as well? Or, it just log the data?
This logger will record the voltage levels that are fed into the input of the logger.  The voltage levels are recorded at a user specified logging interval and the data is viewed in the software on a PC after your recording session is completed.&...
Please, mind that only logged in users can submit questions